Utilizing Coriolis Meters for L.A.C.T. Measurement |
|
By
Michael S. Sulton The Sulton Group/Meter Check P.O. Box 9186 Tulsa, Oklahoma 74107-9186 |
The Crude Oil gathering and transportation industries continue to undergo a period of change in business relationships. New technologies are gaining a significant impact on the manner in which hydrocarbon liquids are measured. The introduction of new types of flow meters has driven the industry to evaluate its' present methods of precision measurement. This paper will address and summarize the History, Theory, Performance, Installation and Calibration of Coriolis Meters for use in L.A.C.T. service.
The driving force for the use of Coriolis technology is the requirement of the industry to reduce the cost of maintenance of traditional flow meters such as Rotary Vane Positive Displacement Meters. The challenge to displace Rotary Vane meters is the ability to provide equal too accuracy with no significant changes to calibration methods or procedures. Both challenges have been successfully overcome by utilizing the non intrusive design of Coriolis flow meters.
"Coriolis force flow meters operate on the principle that inertial forces are generated whenever a particle in rotating body moves relative to that body in a direction towards or away from the axis of rotation of that body" i
In simplest terms, mass flow rate causes the vibrating tubes to twist. The twist of the tubes is proportional to mass flow.
Coriolis force flow meters also utilize vibrating element technology to arrive at density. Once Mass and density are known, volume is calculated.
There exists over a dozen manufactures of Coriolis Flow Meters, these manufactures differ in product offering in sizes and flow rate capability, mounting configurations, how you support the meter, location of other devices, factory testing, published accuracy, field experience.
Typical Performance of Coriolis meters should not be accepted below .15 % accuracy and 20: 1 flow turndown.
Many Coriolis flow meters can readily be adapted to replace existing Rotary Vane meters by simply using a "kit" consisting of welding four elbows and the Coriolis meters mounted in vertical position and a spool piece. Consideration needs to be given to manufactures recommendation on the orientation of the Coriolis. & See figure 3.
In new piping configurations, it is recommend that the Coriolis meter be mounted in the same orientation as a Positive Displacement Meter, typically downstream of a three way divert valve& See figure 4
It is important to note that for volume measurement of Coriolis meters, the calibration procedures should mirror those used on Rotary Vane Positive Displacement Meters.
"Regardless of the prover type or meter type, the same basic proving principle applies"
Know Volume
-------------------- = Meter Factor
Unknown Volume
Meter factor, temperature correction, S&W factor are typically applied in other peripheral devices to calculate gross volume @ standard temperature or net volume.
It is very important to properly apply Coriolis technology to L.A.C.T. design and operation. Proper consideration for good measurement techniques should always take precedence. This technology has proven to be sound and becomes very attractive when a complete cost of ownership is used in the evaluation of Coriolis vs. Rotary Vane meters as the custody transfer device.
i American Petroleum Institute Measurement of Liquids by Coriolis Force Flowmeters
ii Do
you know what would happen if & Frank C arpenter Lake Charles Instrument, Inc.
iii Fundamentals of Meter Proving, Larry Suchy, Mark Vandiver, Meter Check, Inc.
1999,98 Net-Step Media, LLC. All rights
reserved.
Web Design by Net-Step
Media, LLC.
Web Administrator Kirk Harrison
E-mail us your thoughts or suggestions.
Last Update 2-8-00.